Performance of activity recognition versus human observation

VitaMove is medically validated. Perfomance of the VitaMove system was compared to human observance / classification using a protocol of different activities. The following targets for mean of sensitivity and specificity are used for the automatic analysis function in, comparison to manual human observation:

  • > 90 % for Postures like Lying Supine, Prone, Lying left/Right, Standing & Sitting
  • > 80 % for Movements like Moving, Walking, Running, Bicycling.

The following table is a contingency table of all the Body Activities collected over N = 7 subjects. The numbers represent the amount of seconds that the activity was detected.

Man/Auto    Supine    Prone    Leftside    Rightside    Stand    Sit    Move    Walk    Run    Cycle   
Supine 1145 0 0 0 0 25 0 0 0 0
Prone 0 857 14 12 0 0 0 0 0 0
Leftside 15 18 606 0 0 0 0 0 0 0
Rightside 0 66 0 777 0 0 0 0 0 0
Stand 0 0 0 0 674 0 16 0 0 0
Sit 46 0 0 0 0 512 0 0 0 15
Move 0 2 0 0 4 0 232 56 1 90
Walk 0 0 0 0 1 0 17 805 85 36
Run 0 0 0 0 0 0 0 19 643 0
Cycle 0 0 0 0 0 0 0 0 0 195

Sensitivity and specificity

Hereby the actual auto classification results in term of sensitivity and specificity. For all main activities and postures except bicycling, sensitivity and specificity targets are met. Bicycling was ‘real-life’ so contaminated with free-wheeling and such; therefore detection includes ‘Sit’ and ‘Move’ lowering the sensitivity.

Sensitivity Accuracy
Supine 95 100 99
Prone 91 100 98
Leftside 98 99 99
Rightside 98 99 99
Standing 99 100 100
Sitting 95 99 99
Move 88 98 97
Walking 91 98 97
Running 88 100 98
BiCycling 58 100 98